Features for HC-C003 Manufacture Patient monitor price/ 12 inch multiparameter patient monitor
Features for patient monitor price:
12.1'' screen with true color, wide viewing angle, high brightness LCD display.
HR/PR, NIBP, RR, ECG, SpO2, TEMP
Simple and friendly operating display interface.
Rechargeable lithium battery provides convenience for patients' moving.
Playback and browse function for long term waveform and monitor data record.
Optional printing output function, alarm triggers printing.
Auto double alarm with audible and visible signals
Anti-defibrillation, anti-interference from high frequency electric knife
A full-synchronistic lead multi-channel ECG display...
This medical equipment can be used in ICU, CCU and operation room.
Suitable for adult,pediatrics and neonate patients.
Technical Specifications for patient monitor price:
Standard configuration:
HR/PR, NIBP, RR, ECG, SpO2, TEMP, Rechargeable lithium battery
Optional:
Dual TEMP
Dual IBP
ETCO2
Thermal recorder
Wall mount
Trolley
Specification:
ECG
Lead configuration
Standard 3-lead or 5-lead
3-leadRA,LA,LL,Lead method:I,II,III
5-leadRA,LA,LL,RL,V,Lead method:I,II,III,aVR,aVL,aVF,V
Increase ´250,´500,´1000,´2000
HR
Range
Adult 15 ~ 300bpm (beat/minute)
Neonatal babyildren 15 ~ 350 bpm(beat/minute)
Precision±1% or ±1bpm,the larger prevails
Resolution 1 bpm(beat/minute)
Sensitivity> 200 uV (Peak-to-peak value)
Input Impedance> 5 (megohm)
Bandwidth
Diagnostic mode0.05~130Hz
Monitoring Mode 0.5~40Hz
Operation mode1~20Hz
Common Mode rejection Ratio
Diagnostic mode> 90 dB
Monitoring Mode > 100 dB
Operation mode> 100 dB
Pole Polarization Vvoltage Range ±300mV
Pacing Pulse Test
Test pacing pulse in accordance with the following conditions:
Amplitude:±2 mV ~ ±700mV
Width: 0.1ms ~ 2ms
Risetime: 10us~100µs
Pacing pulse inhibition
Amplitude:±2 mV ~ ±700mV
Width: 0.1ms ~ 2ms
Risetime: 10us ~ 100µs
Baseline Recovering Time
After defibrillation< 3 seconds
Signal Range
±8 mV (Peak-to-peak value)
Calibrating Signal
1mV(Peak-to-peak value),precision±5%
ST Segment Measuring Volume
Measuring range: -2.0mV ~ +2.0mV
No definition for other ranges.
RESP
Measuring Method
RA-LL impedance
RESP Impedance Measuring Range
0.3~3Ω
Base Impedance Range
200~4000Ω
Bandwidth
0.1~2.5Hz
RESP Rate
Range
Adult 0~120BrPM
Children and neonatal baby 0~150 BrPM
Resolution 1 BrPM
Precision ±2 BrPM
Asphyxia Alarm
10~40 seconds
NIBP
Measuring Method
Pulse wave oscillometry
Work Mode
Manual/Automatic/STAT
Measuring Interval of Automatic Measuring Mode
1,2,3,4,5,10,15,30,60,90,120,180,240,480 minute(s)
Measuring Time of STAT Mode
5 minutes
PR range
40 - 240 bpm
Overvoltage protection
Adult mode 300 mmHg
Children mode 240 mmHg
Neonatal baby mode150 mmHg
SpO2
Range 0~100%
Resolution1%
Precision 70~100%:±2 DIGIT
0%~69%:no definition given
PR
Range 20~300bpm
Resolution 1bpm Precision ±3bpm