HC-C002 Best price multiparameter patient monitor with 8 inch patient monitor
Features for HC-C002 Best price multiparameter patient monitor with 8 inch patient monitor
8 inch high resolution color TFT LCD display. 5 Standard parameters:ECG, HP, RESP, SPO2, NIBP Built-in lithium battery: 2000mAH Rechargeable Lithium battery Unique accessory management cabinet. Networking capacity. |
RESP
Measuring Method
RA-LL impedance
RESP Impedance Measuring Range 0.3~3Ω
Base Impedance Range 200~4000Ω
Adult 0~120BrPM
Children and neonatal baby 0~150 BrPM
Resolution 1 BrPM
Precision 2 BrPM
Bandwidth 0.1~2.5Hz
RESP Rate Range
Asphyxia Alarm 10~40 seconds
Range
AdultSystolic blood pressure40~270mmHg
Diastolic blood pressure10~215mmHg
Mean blood pressure 20~235mmHg
Children Systolic blood pressure40~200mmHg
Diastolic blood pressure10~150mmHg
Mean blood pressure 20~165mmHg
Static pressure precision3mmHg
Pressure precision:Max. average error: 5mmHg;Max.
standard deviation: 8mmHg
Neonatal baby Systolic blood pressure40~135mmHg
Diastolic blood pressure10~100mmHg
Mean blood pressure 20~110mmHg
Static pressure range 0~300mmHg
Overvoltage protection
Adult mode 300 mmHg
Children mode 240 mmHg
Neonatal baby mode150 mmHg
SpO2
Range 0~100%
Resolution1%
Precision 70~100%:2 DIGIT 0%~69%:no definition given
ECG
Lead configuration
Standard 3-lead or 5-lead
3-leadRA,LA,LL,Lead method:I,II,III
5-leadRA,LA,LL,RL,V,Lead method:I,II,III,aVR,aVL,aVF,V
Increase 250,500,1000,2000
PR
Range 20~300bpm
Resolution1bpm
Precision 3bpm
HR Range
Adult 15 ~ 300bpm (beat/minute)
Neonatal babyildren 15 ~ 350 bpm(beat/minute)
Precision±1% or ±1bpm,the larger prevails
Resolution 1 bpm(beat/minute)
Sensitivity > 200 uV (Peak-to-peak value)
Input Impedance > 5 (megohm)
Bandwidth
Diagnostic mode0.05~130Hz
Monitoring Mode 0.5~40Hz
Operation mode1~20Hz
Common Mode rejection Ratio
Diagnostic mode> 90 dB
Monitoring Mode > 100 dB
Operation mode> 100 dB
Pole Polarization Vvoltage Range 300Mv
Pacing Pulse Test
Test pacing pulse in accordance with the following conditions:
Amplitude:±2 mV ~ ±700mV
Width: 0.1ms ~ 2ms
Risetime: 10us~100μs
Amplitude:±2 mV ~ ±700mV
Width: 0.1ms ~ 2ms
Risetime: 10us ~ 100μs
Pacing pulse inhibition
When pacing analysis switch is on, pacing pulse in accordance with the
following conditions are restrained, but affection against HR calculation.
Baseline Recovering Time
After defibrillation< 3 seconds
Signal Range 8 mV (Peak-to-peak value)
Calibrating Signal
1mV(Peak-to-peak value),precision5%
ST Segment Measuring Volume
Measuring range: -2.0mV ~ +2.0mV
Measuring precision: Ranging -0.8 mV~+0.8mV, measuring error is
±0.02mV or ±10%,.
The larger prevails.