Customization: | Available |
---|---|
Type: | Electrotome |
Application: | Orthopedic, Abdominal, Ent, Neurosurgery, Microsurgery, Anorectal, Urology |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
Product details
Features
Microcomputer control.
Both monopolar and bipolar. Monopolar can be controlled by hand or foot. Bipolar can be controlled by foot only.
Precise power control, can be used for all general surgical procedures.
Touch control panel, three way power output, with different audible and visual indication during the process of outputting.
Automatic performance check during startup and Polar plate test.
Full floating output system protects patient from tissue damage caused by low frequency current.
REM contact quality monitoring system which substantially reduces the risk of burns under the
patient return electrode. When patient plate and skin are not well contacted, the system will automatically
stop the output and sound alarm.
Technical Specifications |
Power: 220V±22V(110V±11V), 50Hz±1Hz, Operating frequency: 360~460kHz Power rating: 1100VA±10% |
Five working modes:
Monopolar Cut
a) pure cut: 1W~300W( Load 500Ω) b) Blend:1W~200W(Load 500Ω)
Monopolar Coag
c)Coag1( Spray Coag):1W~150W( Load 500Ω) d) Coag2(Soft Coag):1W~120W( Load 500Ω)
bipolar coag:1W~50W(70W)( Load 100Ω)
Power Consumption : <=1100VA(Cutting function 300W)
Five working modes:
Monopolar Cut
a) pure cut:1W~300W( Load 500Ω),apply to the operation of any tissue.
b) Blend:1W~200W(Load 500Ω),apply to the surgery of any tissue, proving a good coagulation effect while cutting.
Monopolar Coag
c) Coagulation1(spray Coag):1W~150W( Load 500Ω), apply to ordinary open surgery, deep tissue coagulation layer, has excellent homeostasis effect.
d) Coagulation2(Soft Coag):1W~120W( Load 500Ω) ,suitable for endoscopic and other fine tissue surgery with fast contact coagulation.
Bipolar
e) Bipolar coagulation: 1W~50W ( Load 100Ω), when tissue reach a good coagulation effect,
output power slow down, making a good homeostasis effect.
Power Consumption: <=1100VA (Cutting function 300W, Bipolar coagulation function 50W output).
Packing list
Name |
Quantity |
Mainframe |
1set |
Electrosurgical pencil |
5pieces |
Electrosurgical pad |
10 pieces |
Electrosurgical pad cable |
1piece |
Foot switch |
1set |
Tip clean sponge |
1piece |
Bipolar forceps with cable |
1set |
Fuse |
3pieces |
Power cord |
1piece |
Operating manual |
1book |